Stream Registers
Name Description Address Type Access
STREAM_SETTLING_US
Time in microseconds to allow signals to settle after switching the mux. Does not apply to the 1st channel in the scan list, as that settling is controlled by scan rate (the time from the last channel until the start of the next scan). Default=0. When set to less than 1, automatic settling will be used. The automatic settling behavior varies by device.
4008 FLOAT32 R/W
STREAM_RESOLUTION_INDEX
The resolution index for stream readings. A larger resolution index generally results in lower noise and longer sample times.
4010 UINT32 R/W